NIW-1000 Infant Radiant Warmer

Dedicated for intensive care and resuscitation of new-born and premature baby. So it also called @Resuscitation table@
  • STANDARD COMPOSITION:

    ●Temperature Control Mode: servo control;

    ● Instant Parameter Display: set temp, skin temp, heat power rate;

    ● Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;

    ● Tilting Bed with Rotary Radiant Head;

    ● X-ray Cassette Tray under infant bed, examination light of high illuminance;

    ● Use anti-blast quartz infrared tube as radiant heating source;

     

    TECHNICAL SPECIFICATIONS:    

    Temperature Control Range: 25℃-37℃,

    Temperature sensor accuracy: ≤0.3℃

    Mattress surface temperature uniformity: ≤2.0℃

    Temperature rise time: ≤45mins

    Heat output indicate: 0-100% in 10%

    Radiant box horizon angle adjust range: 0-90°adjustable in step-less

    Bed tilt adjustment: continues adjustable in ±10°

    Mattress Tray Size: 1700px*1350px

    Mattress tray Max. load: 10kg

    Transfusion pole Max. load: 2kg

    Shelf Max. load:2kg

     

    ALARM

    Sensor failure alarm

    Over temperature alarm

    Deviation alarm: ±1℃

    Power failure alarm

     

    ENVIRONMENT TEMPERATURE

    (It is a suggestion that please do not use outside of the allow environment.)

    Operate temperature range18℃~30℃

    Store temperature range-40℃~+55℃

    ENVIRONMENT HUMIDITY

    Operate humidity range30%RH~75%RH

    Store humidity range≤93%RH

    ATMOSPHERIC PRESSURE

    Transport and Store atmospheric pressure range500hPa~1060hPa

    Operate atmospheric pressure range700hPa~1060hPa

     

    STREAM VELOCITY

    Ambit environment stream velocity≤0.3m/s

 
Verification